C.R. BARD, INC. (COVINGTON) -1018233 2.4FR NITINOL TIPLESS BASKET; STONE BASKET Back to Search Results
Model Number 042400
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that in an online survey a physician stated no when they were asked if they were able to successfully remove the renal/ureteral stone.Physician stated they were not able to successfully remove the renal/ureteral stone because impacted stone that did not completely come out in relation to 042400 ¿ skylite tm 2.4 fr.Nitinol tipless basket, 120cm x 12mm.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be due to "basket geometry".It is unknown whether the device had met relevant specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "indications for use: this device is intended for use in endoscopic removal of ureteral and renal stones.Contraindications: none known.Warnings: ¿ some objects may be too large to be removed endoscopically using a retrieval device.The use of fluoroscopy and/or x-ray to determine the size of the object is recommended.Do not use the bard® skylitetm tipless nitinol stone basket if the object is too large to be removed endoscopically, as it may result in patient injury and pain.¿ this is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or may lead to injury, illness or death of a patient.¿ do not attempt to repair, reassemble, or alter the device in any way.¿ after use, this product will be a potential biohazard.Handle and dispose of in accordance with accepted medical practices and with applicable laws and regulations.Follow your institutional guidelines.Caution: objects that are too large to be removed through the scope channel will require the scope and basket to be removed simultaneously from the urinary tract.If resistance is encountered during advancement or withdrawal of the device, stop and determine the source of resistance, as continued resistance may damage the device and could result in patient injury.Take action to alleviate the resistance.Precautions: before using, inspect for any breach of packaging to ensure sterility of product.Do not use if breach in sterile barrier is obvious or suspected.Do not allow the device to come in contact with any electrified instruments or laser.Kinks in the sheath will hinder the mechanical operation of the basket, may affect insertion or withdrawal of the basket and has the potential to damage the endoscope¿s instrument channel.Do not allow the device to be directly fired upon by any lithotripsy devices.To do so may result in damage to the device and could result in patient injury.Capture and removal 1.Under direct vision or fluoroscopic guidance, slowly advance the basket tip past the object.2.Open the basket by pushing the thumb slide forward.(refer to figure b).3.Pull the basket backward toward the object while slowly rotating the basket as necessary.4.Once the object has been captured, partially close the basket to secure the object for removal by carefully pulling the thumb slide back.(refer to figure b).5.Slowly remove the basket and stone from the urinary tract.6.If the object is too large, you may need to simultaneously withdraw the basket and the ureteroscope from the urinary system.Directions for disassembly if handle disassembly is desired or required: 1.Squeeze bottom handle half at indicated points and pull down to remove handle bottom.2.Loosen thumbscrew until basket drive wire moves freely.3.Slide sheath and handle assembly over and away from drive wire." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that in an online survey a physician stated no when they were asked if they were able to successfully remove the renal/ureteral stone.Physician stated they were not able to successfully remove the renal/ureteral stone because impacted stone that did not completely come out in relation to 042400 ¿ skylite tm 2.4 fr.Nitinol tipless basket, 120cm x 12mm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.4FR NITINOL TIPLESS BASKET
Common Device Name
STONE BASKET
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17151372
Report Number1018233-2023-04423
Device Sequence Number14873763
Product Code FFL
UDI-Device Identifier00801741111495
UDI-Public(01)00801741111495
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Study
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number042400
Device Catalogue Number042400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2023
Supplement Date Received by Manufacturer08/03/2023
Initial Report FDA Received Date06/19/2023
Supplement Report FDA Received Date08/04/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-