| Model Number |
7207682 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, the screw biorci got broken while screwing inside the tibial bone.The procedure was completed without surgical delay using a smith and nephew back up device.No further complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, the screw biorci got broken while screwing inside the tibial bone.The broken pieces were removed with a grasper and tweezer.The procedure was completed without surgical delay using a smith and nephew back up device in the original drilled bone.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H2: additional information ¿b5, h6: health effect - impact code¿.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the material specifications found that the storage requirements for the material are specified, and the material must comply with provided percentage composition specifications as measured by ash test.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
|
| |
|
Search Alerts/Recalls
|