| Model Number |
72200873 |
| Medical Device Problem Code |
Smoking (1585)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a shoulder scope, the dii controller had a short circuit and created smoke.The procedure was successfully completed with a non-significant surgical delay using a back-up device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h3, h6: the reported device was received for evaluation.A visual inspection was performed on the exterior of product and no physical damage was observed.A functional evaluation revealed a short circuit detected error message upon plugging unit in.Further evaluation revealed a burnt resistor.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with electrical component failure.Factors that could have contributed to the reported event include connecting a hand piece that is still wet from sterilization processing or connecting a shorted out hand piece causing damage to the motor controller pcb.No containment or corrective actions are recommended at this time.
|
| |
|