• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR OMNILINK ELITE PERIPHERAL STENT SYSTEM; PERIPHERAL STENT DELIVERY SYSTEM Back to Search Results
Model Number 1012631-39
Medical Device Problem Codes Difficult to Remove (1528); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional absolute pro device referenced in b5 is filed under separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a 100% stenosed lesion in the left common iliac artery with heavy calcification and heavy tortuosity.An endarterectomy was performed on the left common femoral artery and an 11fr sheath was placed.A 7x100mm absolute pro self-expanding stent (sess) was advanced to the left external iliac artery and deployed.Then, the 8x39mm omnilink stent delivery system (sds) was brought up to the left common iliac artery, the physician then decided not to deploy the omnilink stent and remove undeployed from the patient.However, during removal the omnilink stent became caught on the absolute pro stent and the stent of the omnilink came off the balloon.The delivery system of the omnilink was removed from the patient; however, the dislodge stent floated through absolute pro stent where it became entangled within the distal portion of the absolute pro.Therefore, physician removed both stents from the patient via a small incision in the endarterectomy patch with a snip tool.There was no adverse patient sequela.A 8x100 absolute pro stent in the left external iliac to complete the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
A visual inspection was performed on the returned stent implant.The reported device damaged by another device and difficult to remove could not be tested, as the stent delivery system was not returned for analysis and due to the condition of the returned stent implant.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The investigation determined the reported difficulties and surgical intervention appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17154188
Report Number2024168-2023-06446
Device Sequence Number5773063
Product Code NIO
UDI-Device Identifier08717648178511
UDI-Public08717648178511
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2025
Device Model Number1012631-39
Device Catalogue Number1012631-39
Device Lot Number1111241
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/26/2023
Supplement Date Received by Manufacturer06/28/2023
Initial Report FDA Received Date06/19/2023
Supplement Report FDA Received Date07/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
Patient Weight44 KG
-
-