• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 980 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number 980A1ENNISB
Medical Device Problem Codes Product Quality Problem (1506); Improper Flow or Infusion (2954)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Issue identified during product analysis.Device evaluation summary: medtronic conducted an investigation based on all information received.It was reported during installation that this 980 ventilator had a failure out of box (foob) due to failing the circuit pressure test during the extended self-test (est) and displayed the error message "inspiratory autozero solenoid not operational" and corresponding diagnostic code.  the device was available for evaluation.The service personnel (sp) inspected the ventilator, confirmed the reported issue, and replaced the inspiratory flow module (ifm) printed circuit board assembly (pcba).The ventilator passed all tests and calibrations according to the manufacturer's specifications at the time of service.One ifm pcba was returned for failure investigation.The part was visually inspected and no anomalies were observed.The part was attached to the test ventilator and powered up but failed the extended self test (est) circuit pressure/autozero solenoid test.After a thorough investigation, a faulty autozero solenoid (so1) in the ifm pcba was identified.If an so1 failure were to occur while in use on a patient, the ventilator response would be to enter backup ventilation (buv).A review of the device history record / product history record identified that the assembly/device had no failures which are relevant to the failure reported by the field.The cause of the event was identified as a faulty autozero solenoid (s01) in the ifm pcba.There is an existing previous internal i nvestigation regarding this issue.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it has met all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported during installation that this 980 ventilator had a failure out of box (foob) due to failing the circuit pressure test during the extended self-test (est) and displayed the error message "inspiratory autozero solenoid not operational" and corresponding diagnostic code. the ventilator was not in use on a patient at the time of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
980 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
COVIDIEN
micheal collins rd mervue
galway
EI 
Manufacturer (Section G)
COVIDIEN
micheal collins rd mervue
galway
EI  
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key17155637
Report Number8020893-2023-00375
Device Sequence Number20277013
Product Code CBK
UDI-Device Identifier10884521201897
UDI-Public10884521201897
Combination Product (Y/N)N
Initial Reporter CountryTH
PMA/510(K) Number
K131252
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/19/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number980A1ENNISB
Device Catalogue Number980A1ENNISB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2023
Initial Report FDA Received Date06/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/19/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-