• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER; PROSTHETIC, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Unspecified Infection (1930); Pain (1994)
Date of Event 01/25/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).D10: 00801802805, femoral head sterile product do not resterilize 12/14 taper, lot number 63897966.110031013, vivacit-e dm bearing 28x46mm, lot number 64689634.00625006525, bone scr 6.5x25 self-tap, lot number 65203637.00784301306, femoral stem beaded fullcoat 12/14 neck taper std.Body std.Offset size 13 13 mm distal dia.160 mm length lot number 62928933.00625006520, bone scr 6.5x20 self-tap, lot number 63029075.110031013, vivacit-e dm bearing 28x46mm, lot number 65233773.Multiple mdr reports were filed for this event, please see associated reports: 0002648920-2023-00127, 0001822565-2023-01646, 0002648920-2023-00004 and 0001822565-2023-00133.The device will not be returned for analysis as it¿s location is not known; however, an investigation of the reported event is in progress.Once the investigation is completed, as supplemental medwatch will be submitted h3 other text : device location is unknown.
 
Event or Problem Description
It was reported that approximately 7 weeks after revision procedure, the patient presented with increased pain and elevated inflammatory markers.After the head and bearing were exchanged, the infection continued without healing requiring a full 2-stage revision.There is no additional information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.A review of the device history records identified no related deviations or anomalies during manufacturing.Medical records were not provided.Upon further review, the root cause of the reported event was determined to be unrelated to the implanted zimmer biomet device; therefore, a review of the device history records will not be performed.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
There is no update to the prior event description provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD STERILE PRODUCT DO NOT RESTERILIZE 12/14 TAPER
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key17157238
Report Number0002648920-2023-00128
Device Sequence Number3893006
Product Code LPH
UDI-Device Identifier00889024144644
UDI-Public(01)00889024144644(17)250531(10)63079295
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
K953337
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/31/2025
Device Model NumberN/A
Device Catalogue Number00801802805
Device Lot Number63079295
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2023
Supplement Date Received by Manufacturer08/17/2023
Initial Report FDA Received Date06/19/2023
Supplement Report FDA Received Date08/25/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PLEASE SEE H10.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age68 YR
Patient SexMale
-
-