| Model Number |
21-4590-24 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Radiation Leak (1357)
|
| Health Effect - Clinical Codes |
Neuropathy (1983); Pain (1994); Implant Pain (4561); Unintended Radiation Exposure (4565)
|
| Date of Event |
05/23/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the 8th neoadjuvant chemotherapy for right breast cancer was completed.During the removal of the butterfly wing needle by heparin saline seal, the bolus injection was not smooth, and the patient felt tingling pain and discomfort in the neck, and the bolus injection was suspended.Contacted the interventional department to perform a digital subtraction angiograph (dsa) to find leakage near the neck.Reported to the doctor in charge and removed the catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: a review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.
|
| |
|