• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL DELTEC PORT-A-CATH IMPLANTABLE VENOUS ACCESS; PORT AND CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL DELTEC PORT-A-CATH IMPLANTABLE VENOUS ACCESS; PORT AND CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAV Back to Search Results
Model Number 21-4590-24
Medical Device Problem Codes Fluid/Blood Leak (1250); Radiation Leak (1357)
Health Effect - Clinical Codes Neuropathy (1983); Pain (1994); Implant Pain (4561); Unintended Radiation Exposure (4565)
Date of Event 05/23/2023
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the 8th neoadjuvant chemotherapy for right breast cancer was completed.During the removal of the butterfly wing needle by heparin saline seal, the bolus injection was not smooth, and the patient felt tingling pain and discomfort in the neck, and the bolus injection was suspended.Contacted the interventional department to perform a digital subtraction angiograph (dsa) to find leakage near the neck.Reported to the doctor in charge and removed the catheter.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
 
Additional Manufacturer Narrative
Device evaluation: a review of the device history record (dhr) shows there were no observations or nonconformities recorded during manufacture to suggest an issue of this nature would occur with this lot of products.No product sample was received; therefore, visual and functional testing could not be performed.The reported issue could not be confirmed.If the product is returned, the manufacturer will reopen this complaint for further investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTEC PORT-A-CATH IMPLANTABLE VENOUS ACCESS
Common Device Name
PORT AND CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAV
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key17157806
Report Number3012307300-2023-06284
Device Sequence Number6021177
Product Code LJT
UDI-Device Identifier10610586021541
UDI-Public10610586021541
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K961580
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number21-4590-24
Device Catalogue Number21-4590-24
Device Lot Number4227823
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/31/2023
Supplement Date Received by Manufacturer06/29/2023
Initial Report FDA Received Date06/19/2023
Supplement Report FDA Received Date07/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/12/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR