• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALPHA OMEGA ENGINEERING LTD. NEURONAV; LNTRAOPERATIVE NEUROPHYSIOLOGICAL RECORDING AND STIMULATING DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALPHA OMEGA ENGINEERING LTD. NEURONAV; LNTRAOPERATIVE NEUROPHYSIOLOGICAL RECORDING AND STIMULATING DEVICE Back to Search Results
Model Number 14
Medical Device Problem Code Disconnection (1171)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The neuronav "system" sn (b)(6) was supplied to the (b)(6) hospital in 2017, it left alpha omega's facility on august 14th 2017.This system last passed preventive "maintenance" on october 18th 2021 before this malfunction, prior to that it passed preventive "maintenance" each year.Following this malfunction, this neuronav main unit was repaired and upgraded to higher device version at alpha omega facility, and shipped back to usa on dec 20, 2022.The system was shipped back to the hospital on jan 31, 2023.Following "receiving" the unit, the hospital "received" case support on april 17th 2023 by alpha omega representative, and reported "successful" procedure.
 
Event or Problem Description
The neuronav hdd laptop failed, so the system laptop failed to boot, an alternative neurosmart system from another hospital was borrowed, so the hospital proceeded with the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURONAV
Common Device Name
LNTRAOPERATIVE NEUROPHYSIOLOGICAL RECORDING AND STIMULATING DEVICE
Manufacturer (Section D)
ALPHA OMEGA ENGINEERING LTD.
hamerkava st.6, tsiporit indus
nof hagalil (nazareth illit), 17890 62
IS  1789062
MDR Report Key17161253
Report Number3001610756-2023-00001
Device Sequence Number5577817
Product Code GZL
UDI-Device Identifier07290014954052
UDI-Public07290014954052
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number14
Device Catalogue Number700-006002-00
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2022
Is the Reporter a Health Professional? Yes
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/20/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/01/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
-
-