Medtronic received information that prior to use during priming, the customer reported one of the custom pack's sample line ports leaked and they noticed it was broken.The device was replaced.There was no patient involvement so no adverse effect occurred.Additional information: the perfusionist provided the sample line from the pack, the remainder of the pack was discarded.The same leak did not occur in other packs.The leak originated from the component.No visible air was noticed in the system/tubing.
|
|
Product analysis # (b)(4): visual analysis: visual inspection shows evidence of some cracks in the luer connection.Performance analysis: pressure integrity testing was performed at 0.5 lpm with 23 psi, (1189 mmhg) of backpressure for 10 min.During the pressure integrity test there was a leak observed at the luer connection.Conclusion: reason for return was confirmed.Unit will be held in brooklyn park.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|