SYNTHES GMBH CHUCK WITH KEY/REAM SPEED FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
|
Back to Search Results |
|
| Model Number |
05.001.207 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/09/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation.The device was evaluated.And it was determined, that the reported condition was not confirmed.Therefore, the assignable root cause was not determined.However, during evaluation, it was determined, that the device was frozen, the chuck seized, the etching was worn, it was difficult to assemble/disassemble and the etching was illegible.The device also failed pretests for marking and labeling, drill chuck with key, maximum range of tool coupling, minimum range of tool coupling and pins length of handpiece coupling.The assignable root cause was traced to component failure, due to normal wear.Udi#: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from switzerland, that during service and evaluation.It was determined, that the chuck with key device was frozen, the chuck seized, the etching was worn.It was difficult to assemble/disassemble.And the etching was illegible.The device also failed pretests for marking and labeling, drill chuck with key, maximum range of tool coupling, minimum range of tool coupling and pins length of handpiece coupling.It was noted in the service order, that the device was not working.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|