| Model Number |
VTICMO12.6 |
| Medical Device Problem Codes |
Inadequacy of Device Shape and/or Size (1583); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2001
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm, vticmo12.6, -14.5/3.0/086 (sphere/cylinder/axis), implantable collamer lens was implanted into the patient's right eye (od) on (b)(6) 2023.Low vault with rotation was observed, the lens was repositioned on (b)(6) 2023, but did not resolve the problem.Lens remain implanted.Cause of the event is reported as other, device did not fail to perform as intended, small size lens with rotation and changing astigmatism.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
A4 - unk.A5 - unk.A6 - unk.H6 - work order search: no similar complaint type events were reported for units within the same lot.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
B5 - lens explanted.H3: device evaluation - lens was returned in microcentrifuge vial dry with residue/debris on product.Visual inspection found optic torn and haptic broken.Dimensional inspection found the lens to be within specification.(b)(4).
|
| |
|