Model Number 393299
Medical Device Problem Codes Failure to Advance (2524); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2023
Type of Reportable Event Malfunction
Event or Problem Description
The pantera pro compliant balloon catheter was selected for treatment of a severely calcified lesion with 95 percent stenosis degree.When pushing the balloon catheter to the lesion site, the proximal end of the catheter was bent.The device could not pass through the calcified lesion.
 
Additional Manufacturer Narrative
The returned product was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The instrument was returned without its packaging.The hypotube is severely kinked at two locations 100 cm proximal to the tip and 2 mm distal to the kink protector.The balloon has been inflated and was returned in a partially deflated state.The remaining instrument shows no damage or irregularities.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause could be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA PRO 1.5/15
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key17163799
Report Number1028232-2023-03086
Device Sequence Number12198631
Product Code LOX
UDI-Device Identifier07640130437036
UDI-Public(01)07640130437036(17)2509
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K160985
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/03/2025
Device Model Number393299
Device Catalogue NumberSEE MODEL NO.
Device Lot Number07224201
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/19/2023
Supplement Date Received by Manufacturer10/02/2023
Initial Report FDA Received Date06/20/2023
Supplement Report FDA Received Date10/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-