| Model Number |
393299 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The pantera pro compliant balloon catheter was selected for treatment of a severely calcified lesion with 95 percent stenosis degree.When pushing the balloon catheter to the lesion site, the proximal end of the catheter was bent.The device could not pass through the calcified lesion.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned product was subjected to a detailed technical analysis and the corresponding product release documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The instrument was returned without its packaging.The hypotube is severely kinked at two locations 100 cm proximal to the tip and 2 mm distal to the kink protector.The balloon has been inflated and was returned in a partially deflated state.The remaining instrument shows no damage or irregularities.Review of the product release documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.Based on the conducted investigations, no material or manufacturing related root cause could be identified.
|
| |
|