| Catalog Number |
60-05-60 |
| Medical Device Problem Code |
Difficult to Open or Close (2921)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/23/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A.1.-a.5.Patient information were not provided.H10: livanova deutschland manufactures the s5 erc tubing clamp.The incident occurred in united states.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
Livanova deutschland has received a report that flow dropped with electronic remote clamp activated by itself during a procedure there is no report of any patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
No additional information has been provided by customer and no request of service activity has been requested.Consequently, it can be assumed that the issue was a one-off event never reoccurred.Complaints database analysis revealed no similar event since unit installation in 2019.Considering the level of information provided, a specific root cause cannot be determined.However, based on similar events reported by other customers the most likely root cause of the event are the following: (i) defective internal board; (ii) faulty communication between the mentioned erc and the related s5 console and/or (iii) user perception ( erc clamp correctly closed following an alarm condition).The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
| |
|
Event or Problem Description
|
|
See intial report.
|
| |
|