EXACTECH, INC. EQUINOXE; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
|
Back to Search Results |
|
| Model Number |
EQUINOXE CAGE GLENOID LARGE, BETA |
| Medical Device Problem Codes |
Fracture (1260); Naturally Worn (2988); Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Insufficient Information (4580); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
05/30/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a 58-year-old male patient, initial right shoulder implanted on (b)(6) 2013, underwent a revision procedure, and was converted to an rsa on (b)(6) 2023, approximately 9 years 6 months post the initial procedure.The surgeon had to revise the glenoid due to loosening and wear.Device breakage was indicated, but no parts or pieces fell into the patient.All fragments were removed.There were no surgical delays reported.No patient history, comorbidities, or x-rays were able to be provided.The patient was last known to be in stable condition following the event.The device is available for return.
|
| |
|
Additional Manufacturer Narrative
|
|
D10: concomitants: 129757, 620-00-02 - platelet rich plasma kit with spray tips.2747500, 300-01-15 - equinoxe, humeral stem primary, press fit 15mm.2774582, 310-01-53 - equinoxe, humeral head short, 53mm (beta).2793267, 300-10-45 - equinoxe replicator plate 4.5mm o/s.2851510, 300-20-02 - equinox square torque define screw drive kit.A20129032, 620-12-02 - accelerate prp 60 ml & acd-a.Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to correct the following field that was entered erroneously in the initial report: g3: 31-may-2023.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h6 - health effect/clinical code, component code additional information: d9 - device receipt h3 - the revision reported was likely the result of extensive eccentric wear of the glenoid component, which led to disassociation of the center cage from the polyethylene body and loosening of the glenoid component from the glenoid bone.Patient conditions, including patient weight and a possible rotator cuff deficiency/tear, likely contributed to the loosening, and the wear and damage observed on the returned device.H6 - medical device problem codes, investigation codes.
|
| |
|