MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® TAG® THORACIC ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number TGT2810
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581); Clinical Endoleak (4843)
Date of Event 06/14/2011
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2011, the patient underwent an endovascular treatment of a retrograde type a dissection and ulp using three gore® tag® thoracic endoprostheses.After all devices were implanted, a proximal type i endoleak was observed.The physician decided to monitor this endoleak.On an unknown date, a follow-up exam revealed the proximal type i endoleak and the false lumen enlargement.No new entry tear was found in false lumen.It was unknown if there was persistent false lumen perfusion.On (b)(6) 2023, a reintervention was performed.A gore® tag® conformable thoracic stent graft with active control system was implanted proximally.The endoleak was resolved.In addition, it was also reported that physician did not report additional information on cause of the type 1 endoleak and the false lumen enlargement occurred due to the type 1 endoleak.
 
Additional Manufacturer Narrative
H6: code e2121-used to capture false lumen enlargement.H6: code c21 - results pending completion of product history review.H6: code c20 - the device remains implanted and, therefore, was not available for direct analysis by gore.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Additional Manufacturer Narrative
H6: code c19-a review of the manufacturing records indicated the lot met all pre-release manufacturing specifications.As the device was not accessible, the product history review (phr) review was the extend of the investigation.No device problem was found per review of these records.It should be noted that, per the gore® tag® thoracic endoprosthesis instructions for use (ifu), adverse events that may occur and / or require intervention include, but are not limited to, endoleaks.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® TAG® THORACIC ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL WOODY SPRINGS B/P
3450 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
lillian espinoza
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key17171259
Report Number2017233-2023-04044
Device Sequence Number7154715
Product Code MIH
UDI-Device Identifier00733132613595
UDI-Public00733132613595
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2013
Device Model NumberTGT2810
Device Catalogue NumberTGT2810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2023
Supplement Date Received by Manufacturer06/23/2023
Initial Report FDA Received Date06/21/2023
Supplement Report FDA Received Date06/26/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/04/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age80 YR
Patient SexFemale
-
-