| Catalog Number |
61-2006 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
05/31/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is not available for evaluation.If additional information is received it will be reported on a supplemental report.
|
| |
|
Event or Problem Description
|
|
It was reported there was device and screw issues post op and a revision surgery was performed.During the revision surgery it was discovered the device and screw was loose.
|
| |
|
Event or Problem Description
|
|
It was reported there was device and screw issues post op and a revision surgery was performed.During the revision surgery it was discovered the device and screw was loose.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker¿s clinical consultant and the engineering department reviewed the ct scan together.They confirmed the 2nd screw from the posterior was loose.Therefore, the reported event (screw loosening) could be confirmed.In conclusion, the patient had a trauma and her discomfort had started after that.The surgeon reported screw loosening, infection symptoms, and low bone quality.Additional surgery showed the bone screw became loose after the trauma and the surgeon removed that screw.If further information becomes available, the investigation will be re-evaluated.
|
| |
|