• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION JAGWIRE; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION JAGWIRE; ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Model Number M00556611
Medical Device Problem Codes Peeled/Delaminated (1454); Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/25/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that a jagwire was used in the sigmoid colon during a colon stent implantation procedure performed on (b)(6) 2023.During the procedure, the guidewire was peeled and detached.The procedure was completed with a new jagwire guidewire.There were no patient complications reported as a result of this event.This event has been deemed a reportable event based on the investigation finding of the core wire broken.Please see block h10 for full investigation details.
 
Additional Manufacturer Narrative
Block h6 (device codes): imdrf device code a0501 captures the reportable event of core wire broken.Block h10: the returned jagwire was analyzed.Upon visual assessment, it was observed that the jagwire ptfe coating area was peeled and accordioned.In addition, the core wire is separated into two sections.Based on the condition of the returned device, engineers determined that the problem observed, is most likely due to device manipulation during the procedure.An excess of force was applied to the device such as during the guidewire insertion through another device or the interaction with the scope, causing the peeled section on the device coating as well as the core wire to be separated into two sections.Boston scientific has determined the most probable cause of this complaint is adverse event related to procedure.It is most likely that the adverse event occurred during the procedure and the device had no influence on the event which led to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAGWIRE
Common Device Name
ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17172173
Report Number3005099803-2023-03267
Device Sequence Number12520817
Product Code OCY
UDI-Device Identifier08714729172161
UDI-Public08714729172161
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K141820
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/14/2026
Device Model NumberM00556611
Device Catalogue Number5661
Device Lot Number0031059970
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/16/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/29/2023
Initial Report FDA Received Date06/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-