• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RFB, MDU, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RFB, MDU, HAND CNTRL, PWRMX EL; SAW, POWERED, AND ACCESSORIES Back to Search Results
Model Number 72200616F
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/05/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference case (b)(4).H3, h6: the reported device was received for evaluation.A visual inspection revealed a discolored cable.Discolored cable is a cosmetic issue that does not affect device functionality and it is not related to the alleged failure.A functional evaluation revealed a blade stall error and overheating of the housing.Further evaluation revealed the drive fork was stiff when rotated.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with a mechanical component failure.Factors that can contribute to overheating include a blade stall condition that will result in increased current draw from the control unit which will heat the motor and hand piece housing.This can be the result of gearbox corrosion caused by cleaning and sterilization methods and the chemicals involved.No containment or corrective actions are recommended at this time.
 
Event or Problem Description
It was reported that during a shoulder arthroscopy, the motor drive unit was no longer working, the blade did not turn.The procedure was completed with a non-significant surgical delay using a back-up device.No further complications were reported.As per results of investigation, overheating of the housing was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RFB, MDU, HAND CNTRL, PWRMX EL
Common Device Name
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key17172312
Report Number1643264-2023-00255
Device Sequence Number7032865
Product Code HAB
UDI-Device Identifier03596010601209
UDI-Public03596010601209
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number72200616F
Device Catalogue Number72200616F
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2023
Initial Report FDA Received Date06/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-