| Catalog Number |
2C4040 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/11/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It reported that there was hair in the tubing of a cysto/bladder irrigation set.The issue was identified during set-up.There was no patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
G1: device manufacturer address 1: 600 mts.Oeste de entrada, principal ave.Las americas, parque industrial.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the device was received for evaluation.A visual inspection was done which observed particulate matter (pm) on the tubing (not inside the tube).The particulate has the following characteristics: black and solid state, and according to laboratory analysis it was identified as a hair.The reported condition was verified.The cause of the condition was due to a manufacturing issue related to human production.Since the pm was found to be outside the fluid path, this issue is not likely to cause or contribute to a death or serious injury.Pm outside of the fluid path does not impact the product delivered to the patient.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|