Model Number IBOT PMD
Medical Device Problem Code Off-Label Use (1494)
Health Effect - Clinical Code Hip Fracture (2349)
Date of Event 05/31/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The action of the user and spouse was an attempt to rotate the clusters in order to switch the position of the two sets of drive wheels and thus try to ensure even tire wear.Stair mode is only intended for use on stairs.Per the user manual, "the device does not balance itself during stair climbing".This is covered in the certificate training program for the user and the assistant during stair climbing.If stair mode is initiated on level ground without steps, and the device is pulled backward, the device tipping and falling over is expected behavior.The initial report from the user stated his spouse only sustained a bruised hip.Upon further discussion with the user five days after the initial report, he clarified that his spouse broke her pelvis as a result of the event.Mobius retrieved the logs from the device.No device malfunction was indicated in the logs.Assessment of the logs corroborated the user's account that a fall had occurred, and indicated that the fall was due to user error.
 
Event or Problem Description
User initiated stair mode on level ground without steps and had his spouse pull back on the device.This resulted in the device falling over backward onto her, breaking her pelvis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IBOT PMD
Common Device Name
IBOT
Manufacturer (Section D)
MOBIUS MOBILITY LLC.
540 north commercial st.
suite 310
manchester NH 03101
Manufacturer (Section G)
MOBIUS MOBILITY LLC.
540 n. commercial street
suite 310
manchester NH 03101
Manufacturer Contact
joseph sullivan
540 n. commercial street
suite 310
manchester, NH 03101
6034258703
MDR Report Key17175129
Report Number3014522447-2023-00002
Device Sequence Number19706495
Product Code IMK
UDI-Device Identifier00857584008010
UDI-Public00857584008010
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K210920
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberIBOT PMD
Device Catalogue NumberMMPX-10000-000
Device Lot NumberN/A
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/05/2023
Initial Report FDA Received Date06/21/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/19/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient SexFemale
Patient Weight59 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-