| Model Number |
DUCT OCCLUDER II |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/31/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2023, a a 5-6mm amplatzer duct occluder was selected for implant in a patient patent ductus arteriosus (pda) using a non abbot delivery sheath.During deployment, the physician noted the device had a cobra deformation and the physician retrieved the device.The deformed device was never released from the delivery cable while inside the patient.There was no report of any interaction with cardiac structures during deployment and no report of any angulation/kink noticed with the delivery system.The procedure was terminated.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.The patient was reported as stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity could not be confirmed.The device was returned to abbott for investigation and the device met functional specifications when analyzed under non-physiological conditions.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and that the product met all defined manufacturing specifications.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with regard to manufacture, design, or labeling.
|
| |
|
Search Alerts/Recalls
|