• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI Back to Search Results
Model Number WA64150A
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2023
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported to olympus, that the base of the opening and closing part of the grasping forceps was damaged.The issue was found during a therapeutic laparoscopic procedure.The physician switched to another grasping forceps and the same thing happened.The procedure was completed with a third set of grasping forceps.There were no reports of patient harm.This event is related to two device malfunctions.This report is being submitted for the first occurrence.The second device is being reported on medwatch patient identifier (b)(6).
 
Additional Manufacturer Narrative
The subject device was returned to an olympus repair center for evaluation and the customer¿s reported problem was confirmed.The device evaluation found that the parts for opening and closing the forceps were broken.The review of the device history records showed that the device was manufactured according to valid instructions and met all specifications.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed that it was caused due to the overloading leading to breakage.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAWS "HIQ+", 5 X 330 MM, GRASPING FORCEPS, CROCEOLMI
Common Device Name
GRASPING FORCEPS
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key17178683
Report Number9610773-2023-01710
Device Sequence Number6391909
Product Code HET
UDI-Device Identifier04042761068970
UDI-Public04042761068970
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K944200
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA64150A
Device Catalogue NumberWA64150A
Device Lot Number21702-013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2023
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2023
Supplement Date Received by Manufacturer03/19/2024
Initial Report FDA Received Date06/22/2023
Supplement Report FDA Received Date03/20/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/09/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-