| Model Number |
WA64150A |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/22/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported to olympus, that the base of the opening and closing part of the grasping forceps was damaged.The issue was found during a therapeutic laparoscopic procedure.The physician switched to another grasping forceps and the same thing happened.The procedure was completed with a third set of grasping forceps.There were no reports of patient harm.This event is related to two device malfunctions.This report is being submitted for the first occurrence.The second device is being reported on medwatch patient identifier (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device was returned to an olympus repair center for evaluation and the customer¿s reported problem was confirmed.The device evaluation found that the parts for opening and closing the forceps were broken.The review of the device history records showed that the device was manufactured according to valid instructions and met all specifications.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the event could not be determined although it can be presumed that it was caused due to the overloading leading to breakage.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|