Device evaluation summary: visual inspection of the returned white luer cap shows evidence of a hole in the stem.Pressure integrity testing was performed at 0.5 l/pm with 23 psi, (1189 mmhg) of back pressure for 10 minutes.During the pressure integrity testing there was a leak observed from the center of the cap.Conclusion: complaint is confirmed for leaks.After evaluation the lot of raw material found no abnormalities, the cause of this complaint was determined as supplier issue due the product analysis indicate the cap has a hole in the stem and in our manufacturing process it is not possible to originate a hole like that.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.There were no adverse patient effects as a result of this incidence.Medtronic will continue to monitor for future occurrences and trends.This regulatory report is being submitted as part of a retrospective review and remediation per (b)(4) as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|