| Model Number |
2307-87-003 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Material Deformation (2976); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/06/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during glenoid preparation the surgeon was utilizing the metaglene positioner plate with the metaglene guide pin.The pin was difficult to pass and appeared to be 'catching as it passed through the central hole'.The positioning guide central hole is burred or roughened internally.The pin was able to be passed and the surgery successfully completed.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary the device was not returned to depuy synthes for evaluation, however photo(s) were provided for review.Review of the provided photo(s) confirms that central hole is burred, surface shows the signs of wear and scratches.And as per visuals we can assume that device is unable to assemble with pin.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|