MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15REC W/ 3L RES; BLOOD GAS OXYGENATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX15REC W/ 3L RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX15RE30C
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H6: component code #1: 483 - manifold.Component code #2: 424 - cap.Health effect - impact code: 2645 - no patient involvement.Heatlh effect - clinical code: 4582 - no clinical signs, symptoms or conditions.Medical device problem code: 1069 - break.Investigation findings: 3233 - results pending completion of investigation.Investigation conclusions: 11 - conclusion not yet available.
 
Event or Problem Description
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during setup, the yellow cap came off from the manifold.No patient involvement.Product was not changed out.Procedure completed successfully.
 
Event or Problem Description
The yellow cap came off from the manifold.
 
Additional Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on june 22, 2023.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added exp date); g3 (date received by manufacturer) ; g6 (indication that this is a follow-up report) ; h2 (follow-up due to additional information and device evaluation) ; h3 (device evaluated by manufacturer); h4 (device manufacture date); h6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Type of investigation #1: 10 - testing of actual/suspected device.Type of investigation #2: 11 - testing of device from same lot/batch retained by manufacturer.Type of investigation #2: 3331 - analysis of production records.Investigation findings: 3259 - improper physical structure.Investigation conclusions: 4307 - cause traced to component failure.The returned sample was inspected upon receipt and confirmed to have 2 yellow caps broken off the manifold.A representative retention sample was inspected for damage with no damage noted on the device.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX15REC W/ 3L RES
Common Device Name
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
125 blue ball road
elkton, MD 21921
7346634145
MDR Report Key17185161
Report Number1124841-2023-00151
Device Sequence Number20456731
Product Code DTZ
UDI-Device Identifier00699753450806
UDI-Public(01)00699753450806
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K151389
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2025
Device Model Number3CX*FX15RE30C
Device Catalogue NumberN/A
Device Lot Number2G27
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2023
Supplement Date Received by Manufacturer07/21/2023
Initial Report FDA Received Date06/22/2023
Supplement Report FDA Received Date07/26/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-