| Model Number |
86200 |
| Medical Device Problem Codes |
Unable to Obtain Readings (1516); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ultra ice plus catheter was evaluated by boston scientific.Visual inspection noted kinks were observed in the imaging window assembly.An image and impedance inspection were performed.Impedance testing shows a good unit wave form.During image characterization testing, a good square image appeared in the ilab system.-microscope inspection did not note any damage consistent with saline exposure post-use of the device.Finally, a functional inspection was done and it was observed that the catheter leaks from the imaging window section.Although laboratory analysis was unable to confirm the reported clinical observations, device presented other failure findings.
|
| |
|
Event or Problem Description
|
|
The complaint is reportable due to pi results.It was reported that an ultra ice plus catheter was selected for use.There was no image displayed despite checking it using a saline solution.No patient complications were reported.The device was returned to boston scientific for laboratory analysis.Upon device analysis, kinks were observed in the imaging window assembly and a leak was observed in the imaging window.
|
| |
|
Search Alerts/Recalls
|