• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE, ORAL/NASAL MAGILL; VAP MICROCUFF ENDOTRACHEAL TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE, ORAL/NASAL MAGILL; VAP MICROCUFF ENDOTRACHEAL TUBES Back to Search Results
Model Number 35111
Medical Device Problem Code Separation Problem (4043)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/11/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.A root cause could not be determined.All information reasonably known as of 21 jun 2023 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.H3 other text : device not returned.
 
Event or Problem Description
Avanos medical inc.Received a single report that referenced ten incidences, which were associated with one device, involving the same patient.This is the fifth of ten reports.See 3011270181-2023-00079 for first event.See 3011270181-2023-00080 for second event.See 3011270181-2023-00081 for third event.See 3011270181-2023-00082 for fourth event.See 3011270181-2023-00084 for sixth event.See 3011270181-2023-00085 for seventh event.See 3011270181-2023-00086 for eighth event.See 3011270181-2023-00087 for ninth event.See 3011270181-2023-00088 for tenth event.It was reported that the microcuff vent connector detached from the endotracheal tube.No injury or medical interventions reported.Per additional information received on 07jun2023, the patient was very active and moved a lot.Staff was present for all but one disconnection.The blue top [vent connector] was immediately reconnected.The patient did not have any associated desaturations.Staff stayed in the room at all times due to concern to prevent harm to the patient.The patient now has a tracheostomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROCUFF PEDIATRIC ENDOTRACHEAL TUBE, ORAL/NASAL MAGILL
Common Device Name
VAP MICROCUFF ENDOTRACHEAL TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
COMMAND MEDICAL NICARAGUA SA
km 12.5 carretera norte
parque industrial las mercedes, edificio 16
managua, ni
NU  
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key17185775
Report Number3011270181-2023-00083
Device Sequence Number12324534
Product Code BTR
UDI-Device Identifier00609038351117
UDI-Public00609038351117
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K080821
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number35111
Device Catalogue NumberN/A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2023
Initial Report FDA Received Date06/22/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age7 MO
Patient SexMale
Patient Weight7 KG
-
-