• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number A2040-080
Medical Device Problem Codes Deflation Problem (1149); Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/31/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the procedure was to treat a mildly calcified, mildly tortuous lesion in the common femoral artery (cfa).An ht command 14 guide wire (gw) was advanced to the lesion without resistance, and a 4.0mmx80mmx150cm armada 14 balloon dilatation catheter (bdc) was advanced over the gw without resistance and inflated without issue.During an attempt to deflate the armada balloon, deflation was difficult, negative was held for approximately 30 seconds and the balloon only partially deflated after multiple attempts.An attempt was made to remove the bdc from the anatomy, however the bdc became stuck on the command gw and both devices had to be removed as a single unit.A new guide wire and balloon dilatation catheter were used to successfully complete the procedure.There were no reported adverse patient effects and no clinically significant delays in the procedure.
 
Additional Manufacturer Narrative
Manufacturer's investigation is still pending at this time.Results and conclusions will be provided in the final report.The additional ht command 14 device referenced in b5 are filed under separate medwatch report number.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported difficulty removing the device was confirmed.The reported deflation problem could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported deflation issue; however, the reported difficulty removing the device appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key17186568
Report Number2024168-2023-06632
Device Sequence Number17750706
Product Code DQY
UDI-Device Identifier08717648156151
UDI-Public08717648156151
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K102705
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2026
Device Model NumberA2040-080
Device Catalogue NumberA2040-080
Device Lot Number2081141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2023
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2023
Supplement Date Received by Manufacturer07/17/2023
Initial Report FDA Received Date06/22/2023
Supplement Report FDA Received Date07/31/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age78 YR
Patient SexFemale
Patient Weight56 KG
-
-