MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723NAB
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/29/2013
Type of Reportable Event Death
Additional Manufacturer Narrative
(b)(4).The pump passed the functional test, including the displacement test, rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and dat at 0.0875 inches.The stop (idle) current and run current measurement tests are within specification.The pump also passed self test, off no power alarm test and a21 error test.The following were noted during visual inspection: minor scratched display window, cracked battery tube threads, missing end cap sticker, and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.History download was successful using thds and carelink upload was successful.The pump did not have a battery installed when received.No damage noted on the original battery cap.Please see below when reviewing the history download.There is no data available due to the pump was stored for an extended period without a battery.Unable to list the total insulin delivered on the event date (b)(6) 2013 and the 7 days prior to that date due to no data available.Unable to perform the history review 1 week prior to the event date (b)(6) 2013 in the pump history file due to no data available.The pump passed the functional testing.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
Information received by medtronic indicated that the customer passed away on (b)(6) 2013.The customer¿s blood glucose was unknown.Troubleshooting was not performed.The insulin pump was returned for product analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723NAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*  00777-3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
*   00777-3869
Manufacturer Contact
justin ellis
ceiba norte ind. park #50 road
juncos 00777--386
*   00777-3869
7635265677
MDR Report Key17188536
Report Number2032227-2023-223507
Device Sequence Number12615857
Product Code OYC
UDI-Device Identifier00613994743893
UDI-Public(01)00613994743893
Combination Product (Y/N)N
PMA/510(K) Number
P150019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723NAB
Device Catalogue NumberMMT-723NAB
Device Lot NumberA3723NABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2023
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2023
Initial Report FDA Received Date06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient Age59 YR
Patient SexMale
Patient Weight100 KG
-
-