| Model Number |
EX071503CS |
| Medical Device Problem Codes |
Misfire (2532); Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/30/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H10: as the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.H10: d4 (expiry date: 11/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3: device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the distal stent two to three cm of the stent deployed and then allegedly became stuck.It was further reported that the stent got deployed after using slide option on the side of the handle.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent placement procedure, the distal two to three centimeter of the stent deployed and then allegedly became stuck.It was further reported that the stent got deployed after using slide option on the side of the handle.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the sample was not returned for evaluation.No photos or x-ray images were provided for review which leads to inconclusive evaluation result.Based on the investigation of the provided information, the investigation is closed as inconclusive.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instruction for use state: 'do not hold the silver stent delivery sheath at any time during deployment.Do not constrict the stent delivery sheath during stent deployment.Initiate stent deployment by rotating the thumbwheel in the direction of the arrows while holding the handle in a fixed position.Note: if excessive force is felt during stent deployment, do not force the stent system.Remove the stent system as possible, and replace with a new unit.' regarding accessories and access the instruction for use state: 'gain femoral access at the appropriate site using a 6f (2.0 mm) or larger introducer sheath.Insert a 0.035¿ diameter guidewire.The instruction for use further state: 'predilation of the lesion should be performed using standard techniques.' h10: d4 (expiry date: 11/2023).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|