| Model Number |
51-402-73-09 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Distractor footplate separated from device spanning likely partial osteotomy.Osteotomy was completed, and distractor was replaced with same model device which performed as expected.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed in the lab and there were no indications of material or manufacturing defects.The complaint percentage was calculated, and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.Review of the device history records was not possible due to no lot number being identified.The investigation results conclude that the root cause for this failure could not be determined.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|