• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON GRIESHABER AG GRIESHABER DSP SURGICAL CONTACT LENS; LENS, FUNDUS, HRUBY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON GRIESHABER AG GRIESHABER DSP SURGICAL CONTACT LENS; LENS, FUNDUS, HRUBY, DIAGNOSTIC Back to Search Results
Catalog Number 618.60
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hyperemia (1904)
Date of Event 04/01/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A clinical survey revealed that after vitrectomy surgery a female patient experienced with mild hyperemia.Unknown intervention was performed to treat the patient.Current patient symptoms were resolved.
 
Additional Manufacturer Narrative
The non-conformance investigation (nci) opened due to monitoring activities of complaint numbers in early june 2023 confirmed that this complaint is related to a post market clinical follow-up (pmcf) study.The pmcf study facilitated a questionnaire, which focused on assessing procedural success of the devices under evaluation.A number of negative responses (i.E., adverse events) was received, which were translated into complaints.Due to the nature of the questionnaire, information required to perform a proper complaint investigation was not collected.No lot number was identified within the study and therefore within this complaint, which is why the associated manufacturing documentation could not be reviewed.All product and batch history records are quality reviewed prior to product release, ensuring that all products on the market fulfill the relevant requirements.Since the questionnaire retrospectively collects data, no sample is available to be provided to the responsible site for an in-depth investigation.Not enough information was provided to properly complete an investigation.The root cause of this complaint could not be identified.The exact root cause for the customer¿s reported event is unknown.Therefore, no specific action can be taken.Within the framework of the mentioned nci, an overall risk assessment was performed and concluded that no further actions are required.Additionally, complaints are reviewed and monitored at regular intervals for adverse trends.No additional actions are needed.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIESHABER DSP SURGICAL CONTACT LENS
Common Device Name
LENS, FUNDUS, HRUBY, DIAGNOSTIC
Manufacturer (Section D)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ  8203
Manufacturer (Section G)
ALCON GRIESHABER AG
winkelriedstrasse 52
schaffhausen 8203
SZ   8203
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key17190309
Report Number3003398873-2023-00084
Device Sequence Number5785306
Product Code HJI
UDI-Device Identifier07612717071285
UDI-Public07612717071285
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number618.60
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/30/2023
Supplement Date Received by Manufacturer08/16/2023
Initial Report FDA Received Date06/23/2023
Supplement Report FDA Received Date09/06/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
VOLK SINGLE USE LENSES
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient SexFemale
-
-