| Model Number |
CYF-5A |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for evaluation and the allegation of bending section rubber ruptured, the metal was sticking out was not confirmed.The device evaluation found due to a cut on the bending section cover, water tightness was lost.The bending section cover has lost adhesive and had slack, scratches, and deterioration.The image guided bundle had significant breakages.Due to wear of the angle wire, bending angle in up/down directions did not meet the standard value.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
An olympus senior field service engineer reported on behalf of a customer, the bending section rubber ruptured, the metal was sticking out and there was not sufficient angulation on the oes cystonephrofiberscope.The issue was found during maintenance.There were no reports of patient or user harm associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and correction to g2 (also selected ¿health professional¿ which was inadvertently left off the initial report), h10 (the complaint was confirmed).The customers complaint was confirmed.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause was unable to be identified.Olympus will continue to monitor field performance for this device.
|
| |
|
Search Alerts/Recalls
|