| Catalog Number |
8065751185 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported that the endoilluminator was stuck in trocar during surgery with no patient contact.The surgery was completed after replacing the product with new one.There was no patient harm reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
The opened illuminator returned for evaluation for the report was not within the final lot number reported based on radio frequency identification tag identification; therefore, no sample evaluation was performed.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.Since the sample associated with the reported product and lot number was not received at the manufacturing site, and the device history record review of the lot number provided indicated the product was processed and released according to the product¿s acceptable criteria; the root cause for the customer complaint issue cannot be determined.The returned sample was not within the final lot number based on radio frequency identification tag identification; therefore, specific actions with regard to this complaint cannot be taken.All fiber optic light guide products are 100% visually inspected and light tested for transmittance and image during manufacturing.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|