| Model Number |
ORT200 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that during the set-up phase of a surgical procedure the imris ort200 operating room table would not lock into position after it was unlocked for rotation.This caused a delay in the start time of the procedure and a patient was on the table at the time of the event.The patient was transferred to an alternate non-imris operating room table that was brought into the operating room.The customer reported interruption to workflow and additional time under anesthesia for the patient, and that the case was completed as intended.No further patient or procedural complications were reported.No intra-operative imaging was planned for the case.
|
| |
|
Additional Manufacturer Narrative
|
|
An imris service engineer assessed the operating room table at the customer site and isolated the cause to an acutator component within the table's rotational locking mechanism.The service engineer removed and replaced the actuator component and functionally tested the table after the repair.The actuator was returned to the manufacturer.Internal corrective action has been initiated by the manufacturer to further address this issue.
|
| |
|