| Catalog Number |
08946353190 |
| Medical Device Problem Codes |
High Test Results (2457); Non Reproducible Results (4029)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.
|
| |
|
Event or Problem Description
|
|
The initial reporter complained of a discrepant high result for 1 patient sample tested for elecsys testosterone ii assay (testosterone ii) on a cobas 8000 e 602 module.The initial result was 52.05 nmol/l with a data flag.This result was reported outside of the laboratory.The physician questioned the result as it did not correspond to the patient's clinical picture and requested repeat testing.The sample was repeated twice with results of 8.21 nmol/l and 7.91 nmol/l.The e602 module serial number was (b)(6).
|
| |
|
Additional Manufacturer Narrative
|
|
A specific root cause could not be determined.A general reagent problem could be excluded because this was an isolated non-reproducible event.
|
| |
|
Search Alerts/Recalls
|