ROCHE DIAGNOSTICS ELECSYS TESTOSTERONE II ASSAY; TESTOSTERONE ASSAY Back to Search Results
Catalog Number 08946353190
Medical Device Problem Codes High Test Results (2457); Non Reproducible Results (4029)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.
 
Event or Problem Description
The initial reporter complained of a discrepant high result for 1 patient sample tested for elecsys testosterone ii assay (testosterone ii) on a cobas 8000 e 602 module.The initial result was 52.05 nmol/l with a data flag.This result was reported outside of the laboratory.The physician questioned the result as it did not correspond to the patient's clinical picture and requested repeat testing.The sample was repeated twice with results of 8.21 nmol/l and 7.91 nmol/l.The e602 module serial number was (b)(6).
 
Additional Manufacturer Narrative
A specific root cause could not be determined.A general reagent problem could be excluded because this was an isolated non-reproducible event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TESTOSTERONE II ASSAY
Common Device Name
TESTOSTERONE ASSAY
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
na
indianapolis, IN 46250
3174767531
MDR Report Key17195029
Report Number1823260-2023-02085
Device Sequence Number20657957
Product Code CDZ
UDI-Device Identifier07613336170557
UDI-Public07613336170557
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K211685
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2024
Device Catalogue Number08946353190
Device Lot Number68876501
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/30/2023
Supplement Date Received by Manufacturer07/05/2023
Initial Report FDA Received Date06/23/2023
Supplement Report FDA Received Date07/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-