• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. NEURON MAX 6F 088 LONG SHEATH; DQY Back to Search Results
Catalog Number PNML6F088804
Medical Device Problem Codes Break (1069); Unraveled Material (1664)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/26/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned neuron max confirmed that the catheter was fractured, and the internal coil winds were unraveled at the hub.Evaluation revealed that the device was kinked at the id band and, upon inspection, was confirmed to be the location of the fracture.If the hub of the neuron max is torqued during manipulation of the device within the reported patient¿s tortuous anatomy, damage such as a kink and subsequent fracture may occur.Further evaluation revealed a kink and ovalization on the device.This damage was likely incidental to the reported complaint and likely occurred during packaging for return to penumbra.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a medical procedure using a neuron max 6f 088 long sheath (neuron max).It was noted that the patient¿s anatomy was tortuous.During the procedure, while injecting contrast through the neuron max, the physician noticed the neuron max had unraveled at the hub.Therefore, the neuron max was removed.The procedure was completed using a new neuron max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON MAX 6F 088 LONG SHEATH
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key17195042
Report Number3005168196-2023-00297
Device Sequence Number7262988
Product Code DQY
UDI-Device Identifier00815948022379
UDI-Public815948022379
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K111380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/23/2023
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/22/2026
Device Catalogue NumberPNML6F088804
Device Lot NumberH00002231
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/07/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2023
Initial Report FDA Received Date06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-