Catalog Number RED72KIT
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/24/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Evaluation of the returned red72 confirmed that the catheter was fractured.If the red72 is mishandled during use, damage such as a kink and subsequent fracture may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text: placeholder.
 
Event or Problem Description
The patient was undergoing a thrombectomy procedure in the internal carotid artery (ica) using a penumbra system red 72 reperfusion catheter (red72) and a non-penumbra sheath.During the procedure, after the red72 was advanced in the patient for the first pass and hooked up for aspiration, the physician noticed that the red72 was aspirating air.Subsequently, the red72 was found to be fractured at the hub.Therefore, the red72 was removed.The procedure was completed using a non-penumbra catheter and the same sheath.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM RED72 REPERFUSION CATHETER
Common Device Name
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key17195473
Report Number3005168196-2023-00298
Device Sequence Number7360289
Product Code NRY
UDI-Device Identifier00815948023932
UDI-Public815948023932
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K211654
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/23/2023
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/25/2025
Device Catalogue NumberRED72KIT
Device Lot NumberH00000778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/05/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/31/2023
Initial Report FDA Received Date06/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Date Report Sent to FDA01/01/2005
-
-