|
A field service engineer (fse) conducted a site visit and was able to confirm the problem by reviewing the quality control (qc) printout.Fse tested the diluent pump, wash probes, and substrate dispense.The customer stated the problem started when they changed diluent and did not give the diluent time to stabilize.During troubleshooting, fse noticed the sample arm demonstrated signs of issues relating to the eki board (liquid surface detection board) failure.Fse ordered a new main arm sampler kit, which includes a new eki board.This was installed at a later day.Fse validated the analyzer by running quality control (qc) with results passing within acceptable range.The aia-2000 analyzer is functioning as expected.No further action required by field service at this time.A 13-month complaint history review and service history review through aware date of event was performed for serial number (b)(6).There were no other similar complaints found during the searched period.The st-aia pack cea analyte application manual states the following: evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The st-aia pack ca125 analyte application manual states the following: evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The st-aia pack 27.29 analyte application manual states the following: evaluation of results quality control in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the controls should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The st-aia pack ca-19.9 analyte application manual states the following: evaluation of results: quality control: in order to monitor and evaluate the precision of the analytical performance, it is recommended that commercially available control samples be assayed daily.The minimum recommendations for the frequency of running internal control material are: after calibration, two levels of controls are run in order to accept the calibration curve.The two levels of controls are also repeated after calibration when certain service procedures are performed (e.G.Temperature adjustment, sampling mechanism changes, maintenance of the wash probe or detector lamp adjustment or change).After daily maintenance, at least two levels of the control should be run in order to verify the overall performance of the tosoh aia system analyzers.If one or more control sample value(s) is out of the acceptable range, it will be necessary to investigate the validity of the calibration curve before reporting patient results.Standard laboratory procedures should be followed in accordance with the regulatory agency under which the laboratory operates.The most probable cause was failure of the sample arm due to faulty eki board (liquid surface detection board).
|
|
A customer reported out of range quality control (qc) for ca125, 27.29, ca-19.9, and cea on the aia-2000 analyzer.All values were more than 50 percent different compared to running the same qc on the customers second aia-2000 analyzer.The analyzer is down.A field service engineer was dispatched to address the reported event, which resulted in a delayed reporting of patient samples for alpha-fetoprotein (afp).There is no indication of any intervention or adverse health consequences due to the delay in reporting of patient results.
|