• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX DUODENAL; STENT,METALLIC,EXPANDABLE,DUODENAL Back to Search Results
Model Number M00565020
Medical Device Problem Codes Unsealed Device Packaging (1444); Packaging Problem (3007)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/07/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a020503 captures the reportable event of seal compromised.
 
Event or Problem Description
Note: this report pertains to one of two devices to be used during the same procedure.Refer to manufacturer report # 3005099803-2023-03453 for the associated device information.It was reported to boston scientific corporation on (b)(6) 2023, that two wallflex duodenal stent was to be used to treat a 7 mm duodenal tumor malignant obstruction in the duodenum during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2023.During procedure, it was noted that the inner sterile packaging seal was compromised (the subject of this report).A different size wallflex duodenal stent (the subject of mfr.Report # 3005099803-2023-03453) was opened and the inner packaging seal was also compromised.A non-boston scientific device was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Event or Problem Description
Note: this report pertains to one of two devices to be used during the same procedure.Refer to manufacturer report # 3005099803-2023-03452 and 3005099803-2023-03453 for the associated device information.It was reported to boston scientific corporation on june 07, 2023, that two wallflex duodenal stent was to be used to treat a 7 mm duodenal tumor malignant obstruction in the duodenum during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2023.During procedure, it was noted that the inner sterile packaging seal was compromised (the subject of this report).A different size wallflex duodenal stent (the subject of mfr.Report # 3005099803-2023-03453) was opened and the inner packaging seal was also compromised.A non-boston scientific device was used to complete the procedure.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a020503 captures the reportable event of seal compromised.Block h10: a wallflex enteral stent and delivery system were received for analysis; the packaging was not returned.The stent was received fully covered by the outer sheath and undeployed.Visual examination of the returned device found the stainless steel separated, the outer blue sheath was bent in different sections, and the outer clear sheath was also found kinked.No other problems were noted with the stent and delivery system.The investigation concluded that the observed failures of stent undeployed, stainless steel separated, outer blue sheath bent, and outer clear sheath kinked were likely due to factors encountered during the procedure.It may be that handling of the device and/or the technique used by the physician (force applied), limited the performance of the device and contributed to the observed failures.The reported event of packaging seal compromised was not confirmed.Based on the available information, there is not enough information to confirm the reported event of packaging seal compromised as the packaging was not returned with the device; therefore, a review and analysis of all available information indicated the most probable cause is no problem detected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX DUODENAL
Common Device Name
STENT,METALLIC,EXPANDABLE,DUODENAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key17198328
Report Number3005099803-2023-03452
Device Sequence Number20229092
Product Code MUM
UDI-Device Identifier08714729456490
UDI-Public08714729456490
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K062750
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/09/2023
Device Model NumberM00565020
Device Catalogue Number6502
Device Lot Number0027461281
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 06/07/2023
Supplement Date Received by Manufacturer08/03/2023
Initial Report FDA Received Date06/26/2023
Supplement Report FDA Received Date08/23/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/09/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-