| Model Number |
BM012R |
| Medical Device Problem Code |
Material Fragmentation (1261)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/06/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap inc.That a tc halsey needle holder serr 130mm (part# bm012r) was used during a hand procedure on june 6 2023.According to the complainant the tip of the device had broken off in the patient during the procedure.Additional information received confirmed that the broken piece was able to be retrieved from the patient.No patient complications were reported as a result of this event.The adverse event is filed under aic reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation results: investigator carried out the pictorial documentation visually and microscopically.The needle holder is in a worn condition, one tungsten carbide inlay is broken, the fragment is not present for examination.The fracture surface of shows no abnormalities like blowholes of foreign particle inclusions.Due to the small dimension/diameter of the device, a detailed fracture surface analysis is difficult.Due to that circumstance there are no clear hints regarding arrestlines (hints for fatigue fracture) or clearly signs of a forced fracture.Device history records: due to the fact, that no lot number was provided, a review of the device history records for the complained device is not possible.The current failure rate is within the risk analysis and therefore acceptable; severity was 4(5) and probability was 1(5).Explation and rationale: the cause of the breakage may be damage caused by a fall of the product.The tungsten carbide inlays have the property of being very brittle, which is why they are more sensitive to such forces.Conclusion/preventive measures: based upon the above mentioned investigation results, a definitive root cause for the reported issue could not be established.Based upon the investigation results, a capa is not necessary.
|
| |
|
Event or Problem Description
|
|
Investigation complete.
|
| |
|