ABBOTT MEDICAL REGENT HEART VALVE; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGN-751
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Cardiac Arrest (1762)
Date of Event 02/27/2023
Type of Reportable Event Death
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2023, a 19mm regent heart valve was selected for an implant into the aortic position.During the procedure the holder handle was fully inserted into the orifice at a 90 degree angle.Fragmentation of the obturator element was noted during fixation of the valve.The leaflet to the valve ring broke off.The dislodged leaflet dislodged as one piece and was removed from the patient.There was no resistance felt when rotating the valve.The physician rotated the valve with the valve rotator.No instruments encountered the valve at the time of dislodgement.The patient collapsed intraoperatively and was put on cardiopulmonary bypass.It was noted that the severity of the event was life threatening.A 19mm regent heart valve was implanted as a replacement.On an unknown date the patient passed away.The cause of the death is multi-organ failure.The implanter classified the death as due to the destruction of the device resulted in a 150% increase in aortic clamping time.No additional information provided.
 
Additional Manufacturer Narrative
An event of leaflet dislodgement when attempting to rotate the valve intra-operatively and patient death was reported.Information from field indicated that a replacement 19 mm regent heart valve was implanted and the overall procedure time was significantly increased.A prolonged implant procedure has the potential to result in a higher risk for bleeding and end-organ failure including renal failure.Field indicated that the cause of the death was due to multi-organ failure, the implanter classified the death as due to the destruction of the device resulted in a 150% increase in aortic clamping time.There was no resistance noted when rotating the valve with the valve rotator.A returned device assessment could not be performed as the device was not returned for analysis.Based on the information received, the root cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REGENT HEART VALVE
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL PUERTO RICO, INC. REG#2648612
20 b st caguas west park
caguas 00725
*   00725
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key17200087
Report Number2135147-2023-02750
Device Sequence Number12183544
Product Code LWQ
UDI-Device Identifier05414734005869
UDI-Public05414734005869
Combination Product (Y/N)N
Initial Reporter CountryRS
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number19AGN-751
Device Catalogue Number19AGN-751
Device Lot NumberC00006317
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/31/2023
Supplement Date Received by Manufacturer07/31/2023
Initial Report FDA Received Date06/26/2023
Supplement Report FDA Received Date08/03/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/21/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexFemale
-
-