| Model Number |
1400250-15 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/05/2023
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the procedure was to treat a de novo lesion located in the distal right coronary artery that was both mildly calcified and tortuous and 75% stenosed.The 2.5x15mm nc trek neo balloon was delivered to the lesion for pre-dilatation, but ruptured at 8 atmospheres during the first inflation.A non-abbott balloon was used for the replacement to perform pre-dilatation and a non-abbott stent was deployed.A non-compliant balloon was used for post-dilatation to complete the procedure.There were no reported adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported balloon rupture.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|
| |
|
Search Alerts/Recalls
|