• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION INTERLINK; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION INTERLINK; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 2N3378
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/30/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an interlink system non-dehp i.V.Catheter extension set leaked.When the tubing was clamped and then the clamp removed, a slight deformity appeared at the site that created a weak point in the tubing.The area ruptured during injection of contrast media.The contrast media was cleaned off the equipment and patient.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.A visual inspection was performed which showed a part of the set with the tubing and the slide clamp with a dark fluid (blood) and the tubing seemed to be burst or ruptured.The reported condition was verified, however, the product code is intended to be used with a syringe or cannula and is not intended to be used with a power injector; the tubing of this set is not designed for the capacity of that kind of equipment as it is not high pressure tubing.Therefore, the cause of the condition was determined to be use related as it was reported the tubing was being used with a power injector.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERLINK
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - AIBONITO
rd 721 km 0 3 po box 1389
aibonito 00705
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key17203700
Report Number1416980-2023-03156
Device Sequence Number6666789
Product Code FPA
UDI-Device Identifier00085412004938
UDI-Public(01)00085412004938
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K925126
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2N3378
Device Lot NumberUR18G18056
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/02/2023
Supplement Date Received by Manufacturer07/23/2023
Initial Report FDA Received Date06/26/2023
Supplement Report FDA Received Date07/28/2023
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/21/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CONTRAST MEDIA
Patient SexUnknown
-
-