| Model Number |
3000 |
| Medical Device Problem Codes |
Gradient Increase (1270); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H10: additional manufacturer narrative: surgical/percutaneous intervention is indicated or performed, or harm occurred due to the device, or there is a device malfunction that could cause or contribute to a serious injury.This event is considered a serious injury.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 23mm 3000 aortic valve underwent a valve-in-valve procedure after an implant duration of 8 years, 4 months due to unknown reason.The tavr was performed with a 23mm 9755rsl transcatheter valve.The patient was stable post procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 23mm 3000 aortic valve underwent a valve-in-valve procedure after an implant duration of 8 years, 4 months due to severe stenosis.The tavr was performed with a 23mm 9755rsl transcatheter valve.The patient was stable and transferred to the pacu post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient-related contributing factors.Structural valve deterioration (svd) is the most common reason for bioprosthesis explant and encompasses multiple failure modes, including calcification, noncalcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, non-structural valve dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors and is not usually an indication of a device malfunction.A definitive root cause cannot be conclusively determined; however, patient factors likely caused or contributed.The device history record (dhr) review was completed and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.
|
| |
|