| Model Number |
VTICM512.6 |
| Medical Device Problem Codes |
Off-Label Use (1494); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The reporter indicated that a 12.6mm vticm512.6 implantable collamer lens of -10.0/2.5/102 (sphere/cylinder/axis) was implanted in the right eye (od) of a patient with pre-existing progressive myopia on (b)(6) 2023.On (b)(6) 2023, the lens was exchanged for a longer due to lens rotation.The problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
A4-a6:unk.H6: off-label - pre-existing progressive myopia.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: h3: device evaluation: the lens was returned in liquid in a microcentrifuge vial with residue/debris on the lens.Visual inspection found the haptic broken and residue/debris on the lens.Dimensional inspection found the lens to be within specifications.Claim#: (b)(4).
|
| |
|