• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® CLOSED SYSTEM URINARY DRAINAGE BAG WITH ANTI-REFLUX CHAMBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDIA® CLOSED SYSTEM URINARY DRAINAGE BAG WITH ANTI-REFLUX CHAMBER Back to Search Results
Model Number 802001
Medical Device Problem Code Restricted Flow rate (1248)
Health Effect - Clinical Code Burning Sensation (2146)
Date of Event 06/12/2023
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the bladder would not drain while drain bag connected to foley.The representative had the patient to check drain bag tubing for kinks and recommended hanging bag lower than body and also to check for catheter for blockage.Per additional information received from liberator on 13jun2023, when the patient was using drain bag urine was not releasing from body and patient experience burning and when the patient swaps to leg bag, and was able to release urine.No medical intervention was reported.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the bladder would not drain while drain bag connected to foley.The representative had the patient to check drain bag tubing for kinks and recommended hanging bag lower than body and also to check for catheter for blockage.As per additional information received from liberator on 13jun2023, when the patient was using drain bag urine was not releasing from body and patient experience burning and when the patient swaps to leg bag, and was able to release urine.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿poor interfaces with connections ¿.It was unknown whether the device had met relevant specifications.A device history record review could not be performed without a lot number.The instructions for use were found adequate and states the following: "directions for use: 1.Wash hands.2.Remove protective cap from drainage tube and connect drainage tube adapter to catheter.3.Fasten drainage bag to bedframe near the foot of the bed.Do not allow bag to touch floor.Important: hang drainage tube in a straight fashion from bedside to drainage bag.If drainage bag is placed even with patient¿s hip, coil tubing alongside patient.Tubing should be draped over patient¿s leg.4.Check that urine is draining into bag.5.To empty bag: a.Remove outlet tube from housing.B.Release clamp and empty bag.C.When bag is empty, return clamp to closed position and replace tube in housing.Note: if specimen is required, see directions for obtaining urine sample.6.Wash hands.7.After use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Directions for obtaining urine sample: 1.Kink tubing a minimum of 3 inches below sample sleeve until urine is under sample site.2.Swab surface of puncture site with antiseptic wipe.3.Insert needle (21g or smaller) through center of puncture site.Be careful to avoid puncturing opposite side of sample sleeve.4.Remove desired volume of urine.Release kink to permit flow to drainage bag.5.Send specimen to laboratory." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDIA® CLOSED SYSTEM URINARY DRAINAGE BAG WITH ANTI-REFLUX CHAMBER
Common Device Name
URINARY DRAINAGE BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key17208189
Report Number1018233-2023-04706
Device Sequence Number17538069
Product Code KNX
UDI-Device Identifier00801741040740
UDI-Public(01)00801741040740
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number802001
Device Catalogue Number802001
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2023
Supplement Date Received by Manufacturer08/08/2023
Initial Report FDA Received Date06/27/2023
Supplement Report FDA Received Date08/11/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexMale
-
-