• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN); PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN); PUMP, BREAST, POWERED Back to Search Results
Model Number 101041360
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Code Inflammation (1932)
Date of Event 06/19/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Customer service conducted troubleshooting with the customer including, verifying no milk backup occurred; inspecting transformer for damaged; trying to power device on with transformer.The customer stated during troubleshooting that the pump will shut off randomly.She also stated she doesn't think trying the portable battery pack will power the pump on.Customer service sent out the customer a new pump in style with maxflow pump.The customer was contacted by a complaint handler on multiple occasions, including in writing, to get additional information, with no response as of the date of this report.Based on the results of our internal investigation (reference number (b)(4), it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as 33%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.(b)(4).Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history of mastitis.Riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.However, in this case, the mother¿s mastitis required prompt medical attention for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.
 
Event or Problem Description
On 6/20/2023, the customer alleged to medela llc that her pump in style with maxflow pump was experiencing power issues.The customer also alleged she was diagnosed with mastitis and was prescribed an antibiotic on (b)(6) 2023.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP IN STYLE MANAGED MARKETS EN FR ES (YELLOWFIN)
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer (Section G)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
Manufacturer Contact
tom cuzman
1101 corporate drive
mchenry, IL 60050
8004358316
MDR Report Key17208888
Report Number1419937-2023-00024
Device Sequence Number5684125
Product Code HGX
UDI-Device Identifier00020451401447
UDI-Public020451401447
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K200508
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number101041360
Device Catalogue Number101041360
Device Lot Number0000430688
Was Device Available for Evaluation? No
Distributor Facility Aware Date06/20/2023
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/20/2023
Initial Report FDA Received Date06/27/2023
Date Device Manufactured01/24/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;