| Model Number |
M00562650 |
| Medical Device Problem Codes |
Failure to Cut (2587); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a sensation large oval med stiff snare was used during a gastroscopy procedure performed on an unknown date.During a gastroscopy procedure, a gastroenterologist performed a macro biopsy in the gastric fundus area.A 30 mm polypectomy snare was used for the biopsy.The snare was tested, and it opened and closed ok.After closing the snare, the handle of the device detached from the device's body.The snare was unable to cut.Bleeding started, and a hemostatic maneuver was performed.It was not reported what intervention was required.The patient had to return the next day and was taken to the operating room to undergo another hemostatic maneuver.It was not reported what intervention was required.Note: no further information has been obtained despite good faith efforts.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block e1: (initial reporter email) (b)(6).Block h6: imdrf patient code e0506 captures the reportable event of hemorrhage major.Imdrf impact code f08 captures the reportable event of hospitalization.Imdrf impact code f23 captures the reportable event of unexpected medical intervention.Imdrf device code a050702 captures the reportable event of loop cutting issues.
|
| |
|