| Model Number |
6020-2530 |
| Medical Device Problem Codes |
Degraded (1153); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Codes |
Ambulation or Postural Difficulties (2544); Metal Related Pathology (4530)
|
| Date of Event |
06/01/2023
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
An event regarding disassociation and wear involving an accolade stem was reported.The event was confirmed via device evaluation.Method & results: device evaluation and results: visual inspection of the returned devices indicated that the trunnion of the stem is severely worn.The metal head appears to have damage from the disassociation event and explantation.Clinician review: no medical records were received for review with a clinical consultant.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been one other similar event for the lot referenced conclusion: it was reported that the patient was revised due to disassociation of the head from the stem.The event was confirmed by device evaluation whereby visual inspection of the returned devices indicated that the trunnion of the stem is severely worn.The reported accolade stem was mated with an lfit v40 cocr head.The cocr head has been identified to be within scope of a nc and capa.No further investigation for this event is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's left hip was revised due to trunnion wear and head dissociation.Intra-operatively, black tissue was noted in the patient, and it was observed that the trunnion had dug into (but not perforated) the liner.Rep will provide any pictures and documents available to him from the hospital and surgeon.
|
| |
|
Search Alerts/Recalls
|